Safe distance from facial nerve for bipolar coagulation in parotid surgery-Animal study.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 39479954.
- Also identified by DOI 10.1002/lary.31883.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Currently no data exist on what distance from facial nerve (FN) it is safe to perform bipolar cautery (BC) in parotid surgery, although frequently performed. The degree of damage was measured using continuous intraoperative neuromonitoring (cIONM, NIM™ 3, Medtronic) in 16 Wistar rats. Amplitude drop of at least 50% (A<sub>50</sub>) or a loss of signal (LOS) in the cIONM was defined as harmful; BC was performed in power range 20-60 W. BC ≤30 W did not cause LOS (0/14 nerves). When applying 35 W, A<sub>50</sub> occurred at 4 mm from FN and LOS was noted in 1 of 5 nerves. BC at a power of 40 to 60 W demonstrated LOS in all nerves (12/12) at a 5 mm distance. BC up to 30 W can be safely applied up to 3 mm distance from FN. 40 to 60 W should be avoided and used only at a distance of over 6 mm from FN. NA/animal study. Laryngoscope, 135:1120-1126, 2025.