IlluminOss photodynamic bone stabilization system improves pain and function in the treatment of humeral metastatic disease.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 39615527.
- Also identified by DOI 10.1302/0301-620X.106B12.BJJ-2023-1089.R2 and PMC identifier 12156336.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of the LightFix Trial was to evaluate the clinical outcomes for one year after the treatment of impending and completed pathological fractures of the humerus using the IlluminOss System (IS), and to analyze the performance of this device. A total of 81 patients with an impending or completed pathological fracture were enrolled in a multicentre, open label single cohort study and treated with IS. Inclusion criteria were visual analogue scale (VAS) Pain Scores > 60 mm/100 mm and Mirels' Score ≥ 8. VAS pain, Musculoskeletal Tumor Society (MSTS) Upper Limb Function, and The European Organization for Research and Treatment of Cancer QoL Group Bone Metastases Module (QLQ-BM22) scores were all normalized to 100, and radiographs were obtained at baseline and at 14, 30, 90, 180, and 360 days postoperatively. The mean VAS pain score decreased significantly from 84 (SD 15) to 50 (SD 29), 38 (SD 30), 31 (SD 29), 31 (SD 29), and 21 (SD 23) between the baseline and follow-up times (p < 0.001). The mean MSTS function scores significantly increased from 27 (SD 19) to 52 (SD 22), 60 (23), 67 (SD 23), 72 (SD 26), and 83 (SD 14) (p < 0.001). The pain and functional subscales of the QLQ-BM22 also significantly improved at most times. A total of 12 devices broke, giving an unadjusted device fracture rate of 15%. Stabilization with the IS decreased pain and improved function with consistent results during the first postoperative year. IS is a new, minimally invasive type of internal fixation. The use of the IS alone may be better for impending rather than completed pathological fractures, and may be better in completed fractures if an added plate or more than the usual number of locking screws is required. Caution is warranted regarding its use alone in patients with a completed pathological fracture due to the rate of breakage of the device.
Medical subject headings
- Bone Neoplasms
- Pain Measurement
- Fractures, Spontaneous
Anatomy
- humerus