Method for the detection and quantification of viral contamination during the preparation of gene therapy drugs in a hospital pharmacy.
Where this comes from
- Record sourced from PubMed, PMID 39632065.
- Also identified by DOI 10.1136/oemed-2024-109574.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The objective of the present study was to develop a method for sampling and detecting an adenovirus-derived gene therapy (GT) vector on isolator worksurfaces. We used a quantitative PCR (q-PCR) to detect the viral genome in standard dilutions of pure GT product and extracts of sampled surfaces. We compared three devices for surface sampling (a cotton compress, a cotton swab and a polyester flocked swab) and performed positive control, negative control and induced contamination tests for each. Our results showed that the GT pure product is detected by the q-PCR assay and is amplified throughout the range of dilutions. The mean difference between the expected and measured number of vector particles in the q-PCR assay was 1.27 log. The numbers of particles in the total extracted volume were 4.66×10<sup>8</sup> for the polyester swab (7.8% of the initial quantity), 3.82×10<sup>8</sup> for a cotton compress (6.4%) and 2.88×10<sup>7</sup> for a cotton swab (4.8%). These initial results suggest that viral monitoring of worksurfaces is feasible and will help us to validate the GT product supply chain.
Medical subject headings
- Genetic Therapy
- Adenoviridae
- Drug Contamination
- Genetic Vectors
- Equipment Contamination