Augmented Reality-Based Finger Joint Range of Motion Measurement: Assessment of Reliability and Concurrent Validity.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 39665730.
- Also identified by DOI 10.1016/j.jhsa.2024.10.006.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study aimed to determine the reliability and concurrent validity of finger joint range of motion (ROM) measurement using augmented reality (AR)-based hand tracking in a sample of healthy hands. Additionally, the study aimed to determine which camera view of the hand provided ROM measurements with the highest concurrent validity at each joint. A web application developed for smart devices using Google's MediaPipe Hands framework converted AR-generated hand landmark coordinates from camera feed into ROM angle measurements in real time for all joints. From each of five camera views, we recorded five sets of AR-based flexion and extension measurements at the metacarpophalangeal (MCP), proximal interphalangeal (PIP) and distal interphalangeal (DIP) joints of normal index to small fingers. Test-retest reliability of the five AR-based measurements in each view was evaluated as was concurrent validity of AR-based measurements relative to manual goniometry, considered the reference standard. Given accepted inter-rater reliability of manual goniometry is 10°, we considered AR-based measurements within 10° of goniometry measurements to have "acceptable" concurrent validity. Forty-eight healthy hands (median age 31, 50% left, varying ethnicities) were measured. All joints demonstrated excellent test-retest reliability (intraclass correlation coefficient >0.75) in all views in flexion and ≥2 views in extension. AR-based flexion measurements were within 10˚ of goniometry in ulnar views of the index MCP, PIP, and DIP; the long MCP and PIP; and the ring PIP and DIP. In extension, multiple views at each joint consistently yielded AR-based measurements within 10° of goniometry. AR-based measurement has high concurrent validity and reliability; however, optimal camera views vary joint to joint. Validation in pathologic hands is required. Given its excellent reliability, AR-based measurement has potential for use in monitoring changes in finger ROM after intervention, either by clinicians in-person or by patients performing remote measurements independently.
Medical subject headings
- Range of Motion, Articular
- Finger Joint
- Augmented Reality