Role of the Laryngeal Adductor Reflex in the Prevention of Thermal Injury to the Recurrent Laryngeal Nerve in Thyroid Surgery: A Case Report.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 39692044.
- Also identified by DOI 10.1002/hed.28038.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Intraoperative neuromonitoring (IONM) of the recurrent laryngeal nerve (RLN) aims to detect and prevent iatrogenic damage during thyroid surgery. Mechanisms of injury include traction, heat damage, and nerve transection. Continuous IONM (C-IONM) techniques detect impending damage due to traction and heat related when they are still reversible. We present a 73-year-old woman who underwent a total thyroidectomy due to a goiter. The IONM included intermittent IONM (I-IONM) in combination with C-IONM using the laryngeal adductor reflex (LAR). The intraoperative neurophysiological signals recorded were correlated with the surgical maneuver and the postoperative evaluation. During surgery, a significant decrease in left LAR amplitude was observed, coinciding with the use of the harmonic. After some corrective maneuvers, the intraoperative signals recovered, and the postoperative examination showed adequate VF function. C-IONM of the RLN using the LAR can be a valuable tool for detecting and preventing thermal spread injury during thyroid surgery.
Medical subject headings
- Thyroidectomy
- Recurrent Laryngeal Nerve Injuries
- Intraoperative Neurophysiological Monitoring
- Reflex
- Intraoperative Complications
- Monitoring, Intraoperative
- Burns
- Goiter