A soft thermal sensor for the continuous assessment of flow in vascular access.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 39746935.
- Also identified by DOI 10.1038/s41467-024-54942-3 and PMC identifier 11696732.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
Hemodialysis for chronic kidney disease (CKD) relies on vascular access (VA) devices, such as arteriovenous fistulas (AVF), grafts (AVG), or catheters, to maintain blood flow. Nonetheless, unpredictable progressive vascular stenosis due to neointimal formation or complete occlusion from acute thrombosis remains the primary cause of mature VA failure. Despite emergent surgical intervention efforts, the lack of a reliable early detection tool significantly reduces patient outcomes and survival rates. This study introduces a soft, wearable device that continuously monitors blood flow for early detection of VA failure. Using thermal anemometry, integrated sensors noninvasively measure flow changes in large vessels. Bench testing with AVF and AVG models shows agreement with finite element analysis (FEA) simulations, while human and preclinical swine trials demonstrate the device's sensitivity. Wireless adaptation could enable at-home monitoring, improving detection of VA-related complications and survival in CKD patients.
Medical subject headings
- Renal Dialysis
- Vascular Access Devices