Non-screw glenoid augmentation constructs for shoulder instability with bone loss: A biomechanical assessment of static and elastic cerclage constructs.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 39846026.
- Also identified by DOI 10.1016/j.jor.2024.12.021 and PMC identifier 11748751.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study aims to compare the biomechanical performance of elastic and static suture-based cerclage systems to traditional screw constructs in the setting of modeled glenoid bony augmentation. Biomechanical testing was conducted on polyurethane cellular foam blocks modeling a 20 % glenoid defect repaired with a coracoid graft. Constructs consisted of an elastic suture-based cerclage, static suture-based cerclage, and a two-screw construct. Biomechanical testing was performed on material testing system, using a 7-phase, 100 cycle per phase, 1Hz, sinusoidal cyclic loading protocol, following a stair-step pattern in load control. Failure for cyclic loading was assessed at 0.8 mm linear displacement. The absolute end level for load-to-failure was 7.0 mm. Static suture-based cerclage failed at 5-50 N (Cycles 2 through 4), 2-screw constructs failed at 25-50 N (Cycle 4), and elastic suture-based cerclage failed at 100-200N (Cycles 6 and 7). Elastic cerclage exhibited superior performance compared to static cerclage beginning in Cycle 2 (p = 0.0440) and compared to SOC 2-screw construct beginning in Cycle 4 (p = 0.0118). 2-screw construct exhibited superior stability performance compared to static cerclage beginning in Cycle 3 (p = 0.0001). Elastic cerclage reached failure at 558.141 ± 4.508 N, while 2-screw construct and static cerclage reached failure at 422.009 ± 24.998 N and 366.770 ± 66.653 N, respectively. Elastic cerclage demonstrated superior biomechanical stability in load-at-failure performance to static cerclage (p < 0.0001) and the screw construct (p < 0.0001), while static cerclage demonstrated inferior biomechanical stability to the screw construct (p = 0.0343). This biomechanical study comparing the performance of elastic cable and static suture tape cerclage fixation methods identified that the elastic cable cerclage exhibits a higher load-at-failure and less displacement under repetitive stress. In addition, elastic cable cerclage fixation exhibits greater strength and construct rigidity than traditional metal screw fixation.
Anatomy
- shoulder