Benefits and Accuracy of Real-Time CGM in Children and Adolescents With Type 1 Diabetes Attending Summer Camp: Second Year.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 40073408.
- Also identified by DOI 10.1210/clinem/dgaf160.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In 2023, we employed Dexcom G6 for real-time continuous glucose monitoring (rt-CGM); it showed high usefulness but unsatisfactory accuracy in summer camp (camp) participants with type 1 diabetes. To assess the usefulness, recommendation, and accuracy of a new rt-CGM system in camp participants in 2024. Sensor glucose (SG) concentrations were measured by Dexcom G7 (G7) from 6 days prior to camp. Usefulness and recommendation were assessed by questionnaire. A total of 21 children with type 1 diabetes and 52 healthcare providers (HCPs) participated. CGM-associated metrics were measured 6 days prior to camp (off camp: 6 days) and during camp (on camp: 2 days). G7 was rated highly for its sensor insertion, visualization of SG variation, and convenience during camp. The SG value accuracy scored relatively low. G7's usefulness and recommendation among campers were significantly higher than those of G6, utilized the previous year, and they were both almost equally highly rated by HCPs. The mean absolute relative difference (MARD) was 11.3% (G6: 16.0%) on camp and 11.0% (G6: 19.7%) off camp. The coefficient of variation was 34.4% (G6: 60.3%) on camp and 35.9% (G6: 63.1%) off camp, while the glucose management indicator was 7.44% for G7 (G6: 8.04%). The glycemic management metrics on camp were better than those off camp. Graphical analyses also showed sufficient accuracy. G7 was rated higher in usefulness and recommendation than G6 among participants. G7 also exhibited improved accuracy and was considered a useful rt-CGM tool in a camp setting. Glucose management on camp was significantly improved because of the multi-directional efforts of the HCPs.
Medical subject headings
- Diabetes Mellitus, Type 1
- Blood Glucose Self-Monitoring
- Blood Glucose