Contrast-Enhanced EIT Robustly Tracks Regional Lung Perfusion Compared to Non-Enhanced EIT and Pulmonary CT.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 40106246.
- Also identified by DOI 10.1109/TMI.2025.3552037.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Regional perfusion monitoring, often performed by pulmonary perfusion computed tomography, is vital in intensive care units. Electrical impedance tomography, repeatable and of non-invasive nature, could provide an attractive alternative. This study compares non-enhanced and contrast-enhanced electrical impedance tomography to computed tomography under induced central to peripheral lung perfusion impairments and cardiac output modulation in 11 animals. A new algorithmic framework using multi-compartment modeling and tracer kinetics was developed to improve perfusion estimation. A multi-resolution mixed models analysis shows electrical impedance tomography agrees poorly with computed tomography in static monitoring, with limits of agreement exceeding relative errors of 100%. For trend tracking,contrast-enhancement with 5.85% NaCl yielded concordance rates above 80%, and over 90% for peripheral impairments, emerging as a robust tracker of coarse to fine perfusion changes. Non-enhanced electrical impedance tomography peaked around 60% under central impairment and cardiac modulation, proving to be less reliable.
Medical subject headings
- Electric Impedance
- Lung
- Tomography
- Tomography, X-Ray Computed
- Contrast Media
- Image Processing, Computer-Assisted