A Comparative Analysis of Glycemic Metrics Derived From Three Continuous Glucose Monitoring Systems.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 40198142.
- Also identified by DOI 10.2337/dc25-0129 and PMC identifier 12178612.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study analyzed the differences in continuous glucose monitoring (CGM)-derived metrics among three current-generation systems and evaluated their impact on therapeutic decision-making. Twenty-three participants wore the FreeStyle Libre 3, Dexcom G7, and Medtronic Simplera CGM systems for 14 days in parallel. CGM metrics were calculated for each participant and CGM system separately. The apparent glucose profile was influenced by the used CGM system, resulting in substantially different glycemic metrics among the three systems. Agreement between FreeStyle Libre 3 and Dexcom G7 was higher than with Medtronic Simplera, which showed lower glucose levels, on average. There were marked intraparticipant discrepancies that would have resulted in different therapeutic recommendations. The CGM systems indicated discordant glycemic metrics, which should be considered in diabetes therapy. Different CGM systems should provide the same glucose readings and CGM-derived metrics when used by the same person.