A Prospective, Multicenter Evaluation of Safety and Diagnostic Outcomes With Robotic-Assisted Bronchoscopy: Results of the Transbronchial Biopsy Assisted by Robot Guidance in the Evaluation of Tumors of the Lung (TARGET) Trial.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 40300665.
- Also identified by DOI 10.1016/j.chest.2025.04.022.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
It remains challenging to safely and reliably biopsy peripheral pulmonary lesions (PPLs). Robotic-assisted bronchoscopy (RAB) is gaining adoption for navigation to PPLs. However, evidence from large studies remains limited. What is the clinical safety, navigational success, and diagnostic yield of RAB for biopsy of PPLs in a broad range of patients in a real-world setting? This multicenter, prospective, single-arm study enrolled patients aged > 21 years with 8- to 50-mm lung lesions requiring bronchoscopic diagnosis. The primary end point was the incidence of the following device- or procedure-related events: (1) pneumothorax requiring intervention; (2) bleeding requiring intervention; or (3) respiratory failure. Secondary end points included individual components of the primary end point, procedure time, pneumothoraces, radial probe endobronchial ultrasound confirmation, conversion to an alternative biopsy procedure, complications, and diagnostic yield. Among 715 patients at 21 sites, 679 met study criteria and underwent RAB (mean age 68.7 years; 55.4% female; 86.5% White; 77.5% with current/past tobacco use). Mean (range) lesion size was 20.9 (7.0-63.0) mm; median (interquartile range) distance from the pleural surface was 5 (0-16) mm. Most lesions were solid (n = 587 [86.6%]) and within the outer two-thirds of the lung (n = 593 [87.5%]). The primary end point was observed in 26 (3.8%) patients (19 pneumothorax, 7 bleeding, and 0 respiratory failure). Users reported that RAB reached the lesion in 670 (98.7%) of 679 cases, and lesion location was confirmed with radial probe endobronchial ultrasound in 607 (91.7%) of 662 cases; sampling through the bronchoscope was performed in 675 (99.4%) of 679 cases. Prevalence of malignancy was 64.1% through 12 months. Adjudicated diagnostic yield was 61.6% when calculated with the American Thoracic Society/American College of Chest Physicians (CHEST) definition for strict reporting criteria. Sensitivity for malignancy was 78.8%. This multicenter prospective study of RAB-to our knowledge, the largest to date-showed that RAB-guided sampling of PPLs is safe and compares favorably to results from sizable non-robotic bronchoscopy studies. ClinicalTrials.gov; No.: NCT04182815; URL: www. gov.
Medical subject headings
- Bronchoscopy
- Image-Guided Biopsy
- Lung Neoplasms
- Robotic Surgical Procedures