Reducing variability in apparent operative inclination (AOI) during total hip arthroplasty: comparing the ball in cage mechanical alignment device to a digital inclinometer.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 40369821.
- Also identified by DOI 10.1177/11207000251323235.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
When operating in lateral decubitus apparent operative inclination (AOI) is the angle between the handle of the acetabular cup introducer and the theatre floor. A digital inclinometer has demonstrated optimal control of AOI. This study aims to compare the deviation from a target AOI with a novel autoclavable ball-in-cage mechanical alignment guide (MAG) as compared to a digital inclinometer. 90 primary total hip arthroplasty cases performed in lateral decubitus using a posterior approach were included. The ball-in-cage mechanical alignment guide on a standard Pinnacle cup introducer handle was used to achieve the target AOI of 35° ± 2.5°. Final AOI was measured with a digital inclinometer. These readings were compared to a published historical series of 90 patients using a digital inclinometer. The ball-in-cage MAG mean inclination was 35.0° (range 32.5-37.4°), with 100% falling within target. This compared to a mean inclination of 34.0° (range 27.5-37.5°), with the digital inclinometer where 87.8% fell within target. However, the range of postoperative radiographic inclination for the ball-in-cage MAG was 23.4-53.9° with 31.1% being outside a ±5° target. These results indicate that this simple autoclavable ball-in-cage MAG consistently achieves an intraoperative AOI within target range. In isolation this does not guarantee a postoperative radiographic inclination within target range.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Patient Positioning
Anatomy
- hip