Pipeline shield reduces diffusion-weighted imaging-detected ischemia after intracranial aneurysm treatment compared with pipeline flex: a propensity score-matched retrospective cohort study.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 40404297.
- Also identified by DOI 10.1136/jnis-2025-023229 and PMC identifier 13217023.
- Licence recorded as CC BY-NC.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The Pipeline Flex embolization device with Shield technology (PED Shield, Medtronic, Irvine, CA, USA) is a flow diverter designed to reduce thrombogenicity through phosphorylcholine surface modification. Although in vitro and in vivo studies have demonstrated the thrombogenicity reducing effects of this technology, its effectiveness in real-world clinical practice remains unclear. This study aims to compare the number of post-procedure diffusion-weighted imaging (DWI)-positive lesions between PED Flex(Medtronic, Irvine, CA, USA) and PED Shield to assess the impact of surface modification. This retrospective cohort study included patients with unruptured intracranial aneurysms treated with PED Flex or PED Shield between April 2016 and March 2024 at a single institution. Propensity score matching was performed to control for confounders, and the number of post-procedure DWI-positive lesions was evaluated as the primary outcome. In total, 148 procedures (132 patients, 138 aneurysms) were included, with 68 (46%) treated with PED Flex and 80 (54%) with PED Shield. Propensity score matching resulted in 47 matched pairs. After matching, the median number of post-procedure DWI-positive lesions was nine (interquartile range (IQR): 3-17) in the PED Flex group and three (IQR: 1-6) in the PED Shield group, with a significantly lower number in the PED Shield group (regression coefficient β = -10.70 [95% confidence interval (CI): -16.23 to -5.16), P<0.001). After adjusting for confounders, the PED Shield group had significantly fewer post-procedure DWI-positive lesions than the PED Flex group, suggesting that phosphorylcholine surface modification technology may reduce thrombogenicity in real-world clinical practice.
Medical subject headings
- Intracranial Aneurysm
- Diffusion Magnetic Resonance Imaging
- Embolization, Therapeutic
- Brain Ischemia