Percutaneous coronary intervention (PCI) with new-generation drug-eluting stents (DES) is still associated with risk of target lesion failure (TLF). The biolimus A9-eluting Biomatrix Alpha stent (BES), with biodegradable polymer and thin struts, has not been compared head-to-head with another contemporary DES. This study compared 1-year TLF in BES vs dual therapy sirolimus-eluting Combo stent (DTS) in an all-comer population undergoing PCI. The trial was conducted in the 3 Western Danish Heart centers (Aalborg, Aarhus, and Odense). The primary composite endpoint was 1-year TLF defined as a composite of cardiac death, target lesion myocardial infarction, or target lesion revascularization. The trial was designed as a noninferiority trial with a noninferiority margin of 2.1%. Data were analyzed by intention-to-treat. From August 14, 2019, to March 19, 2023, 3,136 patients were randomized 1:1 to BES (n = 1,566; 1,891 lesions) vs DTS (n = 1,570; 1,878 lesions). In the intention-to-treat analysis, TLF at 1-year follow-up occurred in 65 patients (4.2%) in the BES group and 82 patients (5.2%) in the DTS group: risk difference: -1.07% (upper limit of 1-sided 90% CI: 0.21%), (P for noninferiority = 0.00002); incidence rate ratio: 0.79 (95% CI: 0.57-1.09; P = 0.15). Cardiac death occurred in 18 patients (1.1%) in the BES group and 30 (1.9%) in the DTS group: incidence rate ratio: 0.60 (95% CI: 0.33-1.07; P = 0.08). Target lesion myocardial infarction occurred in 36 (2.3%) in the BES group and 33 (2.1%) in the DTS group: incidence rate ratio: 1.09 (95% CI: 0.68-1.75; P = 0.73). Definite stent thrombosis occurred in 21 patients (1.3%) in the BES group and 9 (0.6%) in the DTS group: incidence rate ratio: 2.33 (95% CI: 1.07-5.11; P = 0.034). BES was noninferior to DTS at 1-year follow-up regarding the primary endpoint of TLF. However, BES was associated with significantly increased risk of definite stent thrombosis. (Combo Stent Versus Biomatrix Alpha Stent [SORT OUT XI] NCT03952273).

Medical subject headings