Foam film vitrification for cryo-EM.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 40615419.
- Also identified by DOI 10.1038/s41467-025-61270-7 and PMC identifier 12227703.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Electron cryomicroscopy (cryo-EM) has revolutionised structural biology, enhancing applicability, size limits and speed. Despite these successes, cryo-EM sample preparation remains a major bottleneck for routinely achieving high-resolution structures through single particle analysis. Challenges such as inconsistent ice thicknesses, air-water interface interactions and preferred particle orientation persist. Here, we introduce a blot-free vitrification method that uses free-standing surfactant-stabilised foam films to address some of these issues. The method achieves uniform ice thicknesses, enables thickness control of the foam film prior to vitrification, and for some specimens enhances orientation distribution efficiency. Furthermore, it reduces particle adsorption to carbon foil on the specimen support. The method simplifies cryo-EM specimen preparation, offering improved control over ice thickness and particle orientation, to help streamline and accelerate structure determination.
Medical subject headings
- Cryoelectron Microscopy
- Vitrification
- Specimen Handling