Energy consumption of ultrasound devices during routine applications and opportunities to save energy and costs.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 40664868.
- Also identified by DOI 10.1007/s00330-025-11822-8 and PMC identifier 12712049.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
To determine the real energy consumption, including the idle and off-mode status of ultrasound devices and their multiparametric modes during daily examinations. In this study, the electrical power consumption from 9 ultrasound devices and 18 probes used in routine medical practice was measured for a full 24 h over a week. The equipment was located in the neurology and cardiology departments of a large hospital, as well as in a radiology medical practice. The devices were manufactured by different companies and varied in age. Short-term measurements recorded the electrical power consumption during patient scans, revealing real power values for the applied modalities, which ranged between a mean of 182.5 W and 415.5 W, depending on the device and application field. Long-term measurements, reflecting the typical working days of these departments, showed total annual electrical power consumption ranging from 344.6 kWh to 1122.6 kWh. Energy-saving possibilities are discussed, highlighting that reducing idle-mode durations can significantly lower electrical costs by saving energy of 3-16%. This study is the first to document in detail the power consumption of different ultrasound devices and modalities, including idle and off-mode statuses, during routine use. Question Which procedures during ultrasound examinations have high power consumption and what is the idle- and off-mode consumption of different devices. Findings Energy consumption can vary greatly depending on the application. Savings potentials of 3-16% could easily be realized in the daily routine. Clinical relevance Exact knowledge of power consumption of a device during off-, idle-, and some often-used active-modes (all are easy and fast to perform) can help to optimize examination scenarios and minimize power consumption and thus energy costs.
Medical subject headings
- Electric Power Supplies