Assessment and correction of the respiratory phase misalignment using different signals in 4-dimensional imaging and free-breathing gated radiotherapy.

Chen, Lekang; Yu, Shutong; Zhao, Qicheng; Yu, Songmao; Yang, Junqian; Xiao, Jie; Yu, Zijian; Ni, Qianxi et al. · Radiother Oncol · 2025

cross_sectional · Level IV

Where this comes from

Abstract

To propose and validate a correction method for the respiratory phase shift of using different motion management systems in the 4-dimensional (4D) imaging and free-breathing gated radiotherapy. Synchronized cone-beam CT (CBCT) and optical surface images (OSI) of 30 patients at two institutions were included. Reference diaphragm-signals were extracted from CBCT projections using Amsterdam-Shroud (AS) method. Principal Component Analysis (PCA) was used to reduce the dimensionality of OSI data. The phase shift between the AS and PCA signals was calculated using the cross-correlation function. The 20 patients in public 4D-Lung dataset were used for external validation. The relative phase shift averaged over the dataset was proposed as the correction factor. Based on two signals, 4DCBCT were reconstructed using Feldkamp-Davis-Kress (FDK) algorithm respectively. Phase discrepancies of 396 ± 106 ms were observed in the clinical patients. The OSI signals consistently lagged behind the diaphragm signals by 12 % ± 3 % (clinical) and 12 % ± 6 % (public) of a respiratory cycle. Using 12 % as a correction factor, the phase shift was reduced to 76 ms (by 81 %) on average for the 30 clinical patients, and the Root Mean Square Error was significantly decreased by 10-23 % for the phase-averaged 4DCBCT. For 20 public cases, the tumor centroid displacement was reduced by 1.5 ± 0.9 mm, and the inter-sessional variation was 2 % ± 1 % of a respiratory cycle for the 10 cases with multiple 4DCT. The significant phase inconsistency using different motion management techniques should be monitored and corrected before navigating 4D imaging and treatment.

Medical subject headings