Photoacoustic detection of propofol in breath gas for monitoring depth of anaesthesia: from bench to bedside.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 40930873.
- Also identified by DOI 10.1016/j.bja.2025.07.080 and PMC identifier 12597363.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
Ensuring adequate depth of i.v. anaesthesia by measuring propofol in breath gas could increase patient safety. Mass spectrometry, representing the reference standard of propofol breath gas measurements, is not feasible in routine clinical practice; hence, a photoacoustic sensor was developed. The photoacoustic sensor quantifies propofol concentration in gas via the sound waves emitted by propofol molecules excited by light of specific wavelength and frequency. We studied the performance of the new sensor in propofol test gas, gas sampling bags filled with breath gas from different patients, and performed real-time measurements in patients undergoing propofol anaesthesia in comparison to ion-molecule reaction mass spectrometry. In test gas, photoacoustic and mass spectrometry correlated with an R<sup>2</sup> of 0.9975 in a range from 2.5 to 60 ppb. In gas sampling bags, propofol could be detected with both methods. Bland-Altman analysis of propofol general anaesthesia over 18 h in 10 patients revealed a mean propofol difference of -0.02 ppb (standard deviation 3.31) between mass spectrometry and photoacoustic measurements in breath gas, ranging from 4 to 47 ppb. Photoacoustic measurement of propofol concentration in breath gas is feasible with high accuracy in clinical applications.
Medical subject headings
- Propofol
- Photoacoustic Techniques
- Anesthetics, Intravenous