Soft, all-in-one, nanomembrane wearable system for advancing neonatal health monitoring in Ethiopia.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 40998977.
- Also identified by DOI 10.1038/s41746-025-01974-8 and PMC identifier 12462490.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
The neonatal period is critical and stressful, particularly in low-resource settings where existing monitoring methods for newborns are sporadic and labor-intensive. There is a pressing need for an affordable, user-friendly device that offers continuous and effective monitoring. Here, we present a soft, all-in-one, wearable system designed for continuous wireless monitoring of neonatal health in low-resource settings. Developed using advanced packaging technologies, this system features a chest-mounted patch and a forehead-mounted pulse oximeter that transmit real-time data to a smartphone app. The integrated device measures clinical-grade parameters, including electrocardiograms, heart rate, respiration rate, temperature, and blood oxygen saturation, showing excellent correlation with hospital-grade tabletop devices and reduced motion artifacts. This pilot study demonstrates a significant improvement over current monitoring methods by providing continuous oversight, ensures critical events are not missed, and automates the process for enhanced neonatal care, resulting in 84% satisfaction by parents interested in using the system at home.