A novel handheld wireless robot for total knee arthroplasty: Early experiences and clinical results.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 41030820.
- Also identified by DOI 10.1002/jeo2.70444 and PMC identifier 12477546.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
The percentage of total knee arthroplasty (TKA) performed robotically (rTKA) is rising exponentially. While this technique has well-established accuracy improvements, there are mixed patient-reported outcome measure (PROM) benefits compared to manual TKA (mTKA). Potential challenges in workflow, restricted implant compatibility to the robot and system cost have limited robotic adoption. A new robot seeks to address these and other rTKA concerns. This was a single surgeon retrospective series of the first 25 rTKA cases performed with a novel, wireless, handheld, open-implant robot and 25 mTKA cases. Data were recorded to observe performance and safety outcomes, including X-ray-based hip-knee-ankle (HKA) angle measurements, adverse events (AEs), range of motion (ROM), and PROMs including pain and knee society score (KSS). Case times for a subset of cases were recorded. rTKA cases were completed with implants not previously used by the surgeon. mTKA cases used implants and instrumentation standard to the surgeon's practice. Patients were tracked up to 3 months. rTKA and mTKA cases both yielded positive clinical results. There were 0/25 HKA angle outliers (>3° from plan) and 1/25 outliers in the rTKA and mTKA cohorts, respectively. No AEs were observed in either group at follow-up. rTKA case times were similar to mTKA by case 5, with rTKA setup under 8 min for most cases. There were no differences in ROM or PROMs between cohorts. rTKA preliminarily showed safe and effective outcomes in early results. Surgeon and staff adoption showed minimal hurdles at this institution, as rTKA case with an implant system new to the surgeon were similar within five cases to mTKA case times using an implant system known to the surgeon, with comparable clinical results. This initial experience suggests the system may offer a more accessible alternative to existing robotic options, though further study is needed. Level IV.