Robotic arm-assisted total hip arthroplasty enables accurate cup orientation and positioning in obese patients with osteoarthritis secondary to developmental dysplasia of the hip: a propensity score-matched comparative study.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 41048192.
- Also identified by DOI 10.1177/11207000251378086.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study evaluated the accuracy of cup placement by the robotic arm-assisted system (Mako) in obese patients with developmental dysplasia of the hip (DDH). This retrospective single-centre study included 923 consecutive hips with primary robotic arm-assisted total hip arthroplasty (THA). After propensity score-matching, 80 hips with DDH each were selected for the obese group and non-obese group. Postoperative cup orientation and positioning were assessed by superimposition of a 3-dimensional cup template, using postoperative computed tomography images with pelvic coordinates matching the preoperative planning. The absolute error of cup inclination and anteversion was comparable between the obese and non-obese groups (1.2 ± 1.0° vs. 1.2 ± 1.0° and 1.4 ± 1.0° vs. 1.6 ± 1.2°, respectively). The absolute error of the transverse, sagittal, and longitudinal axes between the obese and non-obese groups were 1.0 ± 0.9 mm vs. 1.4 ± 1.0 mm, 1.5 ± 1.0 mm vs. 1.3 ± 1.8 mm, and 1.4 ± 1.2 mm vs. 1.3 ± 0.9 mm (<i>p</i> = 0.02, <i>p</i> = 0.89, and <i>p</i> = 0.12, respectively). There was no significant difference in cup orientation or positioning in the obese group, regardless of body mass index. Robotic arm-assisted THA enables accurate cup placement even in obese patients with DDH.
Anatomy
- hip