The Importance of Modeling Press-Fit to Accurately Evaluate Interfacial Micromotion as an Indicator of Primary Stability in Uncemented Arthroplasty.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 41063630.
- Also identified by DOI 10.1002/jor.70081.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Arthroplasty is most often performed to alleviate pain and restore function in patients with end-stage degenerative joint disease. Uncemented implant fixation is increasingly used in total knee and total ankle arthroplasty. Implantation using interference press-fit is a manufacturer-recommended guideline for achieving stable primary fixation in uncemented applications, which is important to prevent long-term implant failure due to aseptic loosening. However, when evaluating implant-bone interfacial mechanics, many studies have not modeled press-fit implantation. This can lead to gross underestimation of primary implant fixation stability, limiting the clinical applicability of findings. The goal of this paper is to highlight the importance of simulating press-fit implantation when evaluating primary orthopedic implant stability using finite element analysis in uncemented arthroplasty. Experiences gained in modeling press-fit implantation in total knee and total ankle arthroplasty by two different active research groups are shared in this context.
Medical subject headings
- Arthroplasty, Replacement, Ankle
- Arthroplasty, Replacement, Knee
Anatomy
- ankle
- knee