Subcellular mass spectrometry imaging of lipids and nucleotides using transmission geometry ambient laser desorption and plasma ionisation.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 41093837.
- Also identified by DOI 10.1038/s41467-025-64604-7 and PMC identifier 12528750.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
Demand for mass spectrometry imaging (MSI) technologies offering subcellular resolution for tissues and cell imaging is rapidly increasing. To accomplish this, efficient analyte ionisation is essential, given the small amounts of sample material in each pixel. Herein, we describe an atmospheric pressure transmission-geometry matrix-assisted laser desorption source equipped with plasma ionisation. By utilising a pre-staining method for sample preparation, lipid signal intensities were enhanced by an order of magnitude compared to conventional matrix-only methods, while serendipitously enabling imaging of numerous nucleotides. The system enables detection of up to 200 lipids and nucleotides in tissues at 1 µm-pixel size while informative MSI data is still obtained down to 250 nm pixel size. The use of sub-micron pixels is shown to discern subcellular features through combinations with fluorescence microscopy. This method provides a powerful tool for cellular and sub-cellular imaging of small molecules from tissues and cells for spatial biology applications.
Medical subject headings
- Lipids
- Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization
- Nucleotides
- Molecular Imaging