Early Failure Risk and Long-Term Outcomes of Advanced Bionics Electrode Devices Post-Recall.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 41104824.
- Also identified by DOI 10.1002/ohn.70048.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To characterize the timing and likelihood of electrode failure in Advanced Bionics (AB) cochlear implants and assess long-term trends following the 2020 recall of HiRes Ultra and Ultra 3D models. Retrospective cohort study. Single tertiary academic medical center. We conducted a retrospective chart review of patients who received AB HiRes Ultra or Ultra 3D cochlear AB implants between January 2014 and February 2020. Patients who experienced device failure requiring revision surgery were identified. Demographics, implant date, failure status, and time to failure were recorded. Kaplan-Meier survival analysis was used to model cumulative failure risk over time. A total of 140 AB implants were placed, with 24 (17.1%) device failures identified. The median time to failure was 3.6 years (range: 0.39-6.0 years). Median follow-up was 5.2 years (range: 0.1-10.7 years). There was no significant association between failure and gender, age, race/ethnicity, or laterality. Kaplan-Meier analysis demonstrated that 95% of failures occurred within the first 5.9 years, and the failure rate decreased to 2.1% per year after 5 years. AB cochlear implant failures primarily occur within the first 5 years postimplantation, with significantly reduced risk beyond this period. These findings support long-term device stability and can inform patient counseling, expectations, and follow-up protocols.
Medical subject headings
- Cochlear Implants
- Prosthesis Failure
- Cochlear Implantation
- Medical Device Recalls