Prospective multicentre single-arm study of an interatrial shunt in heart failure with reduced ejection fraction (SUSTAIN-HF): 1-year clinical and haemodynamic outcomes.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 41107031.
- Also identified by DOI 10.1136/heartjnl-2025-326319.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Evidence for atrial shunt devices in heart failure with reduced ejection fraction (HFrEF) is limited. This study aimed to evaluate the 1-year clinical and haemodynamic outcomes associated with an atrial shunt device in this population. In this prospective, multicentre, single-arm cohort study of an interatrial shunt in heart failure with reduced ejection fraction (SUSTAIN-HF), 120 symptomatic HFrEF patients (left ventricular ejection fraction; LVEF ≤40%) on guideline-directed medical therapy were enrolled across 16 centres in China. The intervention was implantation of the D-Shant atrial shunt device. Prespecified primary outcomes were the changes from baseline to 1 year in 6 min walk distance (6MWD) and New York Heart Association (NYHA) functional class. Secondary outcomes included quality of life (Kansas City Cardiomyopathy Questionnaire; KCCQ), echocardiographic parameters and adjudicated adverse events. Between October 2017 and December 2021, 120 patients were enrolled. At 1-year follow-up, implantation of the shunt device was associated with significant improvements in clinical outcomes. The mean 6MWD increased by 54.1 m (95% CI 45.9 to 62.3; p<0.0001), and the KCCQ overall score was improved by 16.5 points (95% CI 13.9 to 19.1; p<0.0001). The proportion of patients in NYHA class III/IV decreased from 95.0% to 22.9% (p<0.0001). The annualised rate of heart failure hospitalisation was reduced from 3.3 to 0.3 events per patient-year (p<0.0001). LVEF increased by 9.0% (95% CI 6.9% to 11.1%; p<0.0001). The primary safety endpoint, a composite of major adverse cardiac and device-related events at 1 year, occurred in 15.0% of patients, driven primarily by cardiovascular death. In this prospective cohort study of symptomatic HFrEF patients, treatment with the D-Shant atrial shunt device was associated with significant improvements in functional capacity, quality of life and rates of heart failure hospitalisation at 1 year. The single-arm design precludes definitive conclusions on treatment efficacy, underscoring the need for randomised, sham-controlled trials.
Medical subject headings
- Heart Failure
- Stroke Volume
- Hemodynamics
- Ventricular Function, Left
- Heart Atria