BrachyPlan: A fine-grained efficient dose-guided inverse planning strategy for low-dose-rate brachytherapy.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 41197227.
- Also identified by DOI 10.1016/j.media.2025.103852.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Brachytherapy delivers highly conformal doses for malignancies ranging from pancreatic to head-and-neck cancers, yet today's treatment-planning systems still depend on extensive manual manipulation and dose engines of uncertain accuracy. We introduce BrachyPlan, an end-to-end pre-operative framework that automates low-dose-rate (LDR) brachytherapy planning while preserving clinical precision. The system combines: 1) a real-time Monte Carlo dose engine that produces high-fidelity, heterogeneity-aware dose maps, and 2) a dose-guided inverse-planning algorithm that optimally places seeds to meet target coverage and organ-at-risk constraints. In retrospective experiments, BrachyPlan reproduced ground-truth Monte Carlo dose-volume histograms (DVHs) with <5% error across all metrics, achieved V100 >95% for every clinical target, and cut planning time to one-eighth of the manual workflow, with minimal manual intervention and optimal efficiency. To our knowledge, this is the first platform that unites a clinically validated dose calculation algorithm with real-time inverse optimization, delivering both sub-5% dosimetric accuracy and unprecedented efficiency for LDR brachytherapy.
Medical subject headings
- Brachytherapy
- Radiotherapy Planning, Computer-Assisted