Threshold screw insertion torque for carbon fibre-reinforced polyetheretherketone and titanium (Ti-6Al-4V) locking plate constructs.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 41671892.
- Also identified by DOI 10.1016/j.clinbiomech.2026.106769.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Locking plate constructs are integral to modern fracture fixation, particularly in osteoporotic bone. Carbon-fibre reinforced polyetheretherketone (CFR-PEEK) plates offer a stiffness profile closer to cortical bone than titanium, potentially improving the mechanical environment for healing. However, the optimal insertion torque for locking screws in CFR-PEEK plates is unclear. This study investigated the influence of insertion torque on construct performance in CFR-PEEK plates compared to titanium alloy (Ti-6Al-4V) plates. Locking screws (3.5 mm, Ti-6Al-4V) were inserted into Ti-6Al-4V and CFR-PEEK plates at six torque levels (0.5-3.0 Nm). Construct strength was assessed via axial push-out and cantilever bending tests. Video analysis was used to correlate insertion torque with screw rotation. Additional testing was performed on two-screw CFR-PEEK constructs at three torque levels (0.5, 1.5, 2.5 Nm). Insertion torque was linearly related to screw rotation within the tested range (0.5-3.0 Nm). In CFR-PEEK single screw constructs, higher torque (≥2 Nm) improved push-out strength (p < 0.05), while 1.5 Nm yielded the highest cantilever strength (p < 0.05). Ti-6Al-4V constructs showed a similar trend. In two-screw CFR-PEEK constructs, no significant differences were found in performance across torques for either push-out or cantilever testing (p > 0.5). Beyond an initial threshold (∼0.5 Nm), increased insertion torque did not consistently enhance construct performance. These findings indicate that moderate torques (∼1.5 Nm) optimise performance while reducing the risk of implant damage, support the mechanical reliability of CFR-PEEK plates.
Medical subject headings
- Titanium
- Ketones
- Polyethylene Glycols
- Bone Plates
- Bone Screws
- Carbon