MEPs elicited by multidirectional rotational-field TMS show marked differences compared to unidirectional Figure-of-8 and H7 coils.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 41746909.
- Also identified by DOI 10.1371/journal.pone.0343725 and PMC identifier 12944766.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Unidirectional transcranial magnetic stimulation (udTMS; e.g., via Figure-of-8 coil) depolarizes mainly neurons whose axonal orientation aligns with the direction of the induced electric field. A novel dual H-coil (T360°) TMS system (BrainsWayTM, Israel) generates a rotational magnetic field aimed to recruit a larger neuronal population by induction of a multidirectional electric field (rfTMS). This study aimed to comparatively assess the neurophysiological properties of motor evoked potentials (MEPs) elicited from the first dorsal interosseous (FDI) muscle following udTMS (via Figure-of-8 and H7 coils) vs. multidirectional rfTMS. In this study, 10 healthy adult subjects received TMS via the three coil configurations in a random order. The results showed that rfTMS elicited larger MEPs at a lower resting motor threshold (rMT) compared to the unidirectional coils. These findings suggest that rfTMS is likely to recruit larger populations of neurons compared to conventional udTMS coil configurations. This may be advantageous in efforts to enhance motor recovery following brain damage by treatments using TMS.
Medical subject headings
- Transcranial Magnetic Stimulation
- Evoked Potentials, Motor