Subgroup performance of a commercial digital breast tomosynthesis model for breast cancer detection.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 41857004.
- Also identified by DOI 10.1038/s41467-026-70637-3.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
AI models show potential to improve breast cancer screening, however detailed subgroup evaluations to uncover the strengths and weaknesses of models are lacking. This study presents a granular evaluation of a commercial AI model for cancer detection on digital breast tomosynthesis (DBT) on a retrospective cohort of 167,860 screening exams in female patients. Performance in distinguishing screen detected cancers (1,368 exams) from negative exams (166,387 exams) is stratified across demographic, imaging, and pathologic subgroups to identify disparities. The overall AUROC is 0.91 and sensitivity is 0.73 with robust performance across demographics. In-situ cancers (AUROC: 0.85, sensitivity: 0.55), calcifications (AUROC: 0.80, sensitivity: 0.66), and dense breast tissue (AUROC: 0.88, sensitivity: 0.63) are associated with lower performance, while masses (AUROC: 0.93, sensitivity: 0.85) and architectural distortions (AUROC: 0.90, sensitivity: 0.83) are associated with higher performance. These results highlight the need for detailed evaluations and vigilance in adopting new clinical tools.