Autonomous OCT-Guided Robotic System for Vertical Needle Insertion and Big-Bubble Dissection in Corneal Transplantation.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 41945802.
- Also identified by DOI 10.1109/TBME.2026.3681869 and PMC identifier 13137264.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Autonomous surgical robots can improve consistency and safety in clinical outcomes, independent of surgeon expertise. Deep anterior lamellar keratoplasty (DALK), a partial-thickness corneal transplant that separates Descemet's membrane (DM) from the anterior stroma, demands high precision to avoid perforating the DM during pneumodissection. We introduce AUTO-DALK, a real-time optical coherence tomography (OCT) sensing robotic system for autonomous vertical needle insertion into deep stromal layers. Unlike conventional horizontal approaches that risk incomplete dissection or DM perforation, AUTO-DALK employs a sensor-guided vertical trajectory to achieve more uniform pneumodissection. The system integrates an OCT fiber optic sensor with an anatomical segmentation model to enable real-time depth perception and closed-loop needle control. AUTO-DALK was evaluated in both ex vivo and in vivo rabbit models against 1) freehand, 2) OCT-guided manual, and 3) teleoperated robotic insertions. Results show superior performance in needle placement, dissection accuracy, completion time, and big bubble formation, demonstrating the feasibility of OCT-guided autonomous robotic execution for partial-thickness keratoplasty.