The confounding effects of skin colour in photoacoustic imaging.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 42310010.
- Also identified by DOI 10.1038/s41467-026-74380-7.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Skin colour is known to confound optical devices, adversely impacting care for patients with darker skin. Photoacoustic imaging (PAI) combines optics and ultrasound for deep tissue imaging, creating a complex relationship between PAI-derived biomarkers and skin melanin concentration, yet no generalisable bias correction has been demonstrated. Drawing on a healthy volunteer cohort of 42 participants spanning Fitzpatrick types I-VI and vitiligo - the most diverse ever assembled in PAI - we characterise optical and acoustic mechanisms driving skin colour-dependent degradation in image quality and biomarker quantification. Wavelength-dependent melanin absorption causes spectral colouring, dominating at low melanin levels, while epidermal ultrasound backscattering dominates at high melanin levels, producing a non-linear relationship between unmixed sO₂ and skin tone. Leveraging this understanding, we propose a practical spectral colouring correction and adapt a plane-wave reconstruction algorithm to resolve backscattered ultrasound artefacts. Our findings underscore the need for advanced reconstruction methods to enable equitable clinical PAI.