Radiation scatter at fluoroscopy-guided endoscopic procedures: Building towards further optimization.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 42358121.
- Also identified by DOI 10.14309/ajg.0000000000004097.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Exposure to ionizing radiation by endoscopy personnel during fluoroscopy-guided procedures remains a health hazard. We aimed to evaluate a novel lead flap-enhanced artificial intelligence-enabled fluoroscopy (LE-AIF) system to minimize radiation scatter. We conducted prospective study of 218 patients who underwent fluoroscopy-guided endoscopic procedures using conventional AIF (C-AIF) or LE-AIF systems. The primary outcome measure was comparison of radiation scatter to endoscopists. Radiation scatter to endoscopists and anesthesia personnel were significantly lower using LE- AIF. On linear regression analysis, C-AIF and fluoroscopy duration were associated with radiation scatter. LE-AIF significantly reduced radiation scatter at endoscopic procedures.