A new method for on-screen ultrasonographic determination of fetal cardiac axis.

Bork, M D; Egan, J F; Diana, D J; Scorza, W E; Fabbri, E L; Feeney, L D; Campbell, W A · Am J Obstet Gynecol · 1995

prospective_cohort · Level II

Where this comes from

Abstract

Our purpose was to develop an ultrasonographic method of using multiple electronic calipers for on-screen measurement of fetal cardiac axis. Two hundred low-risk patients who were seen for antenatal ultrasonography were studied. Standard biometry, anatomic survey, and echocardiography were performed on all fetuses. By use of a four-chamber view fetal cardiac axis was measured with electronic calipers. A table, derived from a trigonometric formula, was created to convert the caliper measurements to cardiac axis in degrees. The results were compared with a protractor-determined fetal cardiac axis. All echocardiograms had normal results. Fetal cardiac axis (+/- 2 SD) by the ultrasonographic method was 43 degrees (+/- 16 degrees) versus 43 degrees (+/- 14 degrees) by the protractor method. This difference was not significant. An on-screen method to determine fetal cardiac axis by use of multiple calipers is described. It is comparable to a protractor-measured fetal cardiac axis.