A new method for on-screen ultrasonographic determination of fetal cardiac axis.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 7485318.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Our purpose was to develop an ultrasonographic method of using multiple electronic calipers for on-screen measurement of fetal cardiac axis. Two hundred low-risk patients who were seen for antenatal ultrasonography were studied. Standard biometry, anatomic survey, and echocardiography were performed on all fetuses. By use of a four-chamber view fetal cardiac axis was measured with electronic calipers. A table, derived from a trigonometric formula, was created to convert the caliper measurements to cardiac axis in degrees. The results were compared with a protractor-determined fetal cardiac axis. All echocardiograms had normal results. Fetal cardiac axis (+/- 2 SD) by the ultrasonographic method was 43 degrees (+/- 16 degrees) versus 43 degrees (+/- 14 degrees) by the protractor method. This difference was not significant. An on-screen method to determine fetal cardiac axis by use of multiple calipers is described. It is comparable to a protractor-measured fetal cardiac axis.
Medical subject headings
- Fetal Heart
- Ultrasonography, Prenatal