Plaque abrasion and intra-aortic balloon leak.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 7497749.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
STUDY OBJECTIVE: To determine and correct cause of high incidence of intra-aortic balloon leaks (ruptures). DESIGN: Epidemiologic investigation of factors associated with intra-aortic balloon leak, and sequential application of corrective measures evaluated by continued concurrent data collection. SETTING: Thirty-four-bed ICU in 598-bed tertiary care medical center. INTERVENTIONS: Procedure changed to place smaller balloons (34 mL instead of 40 mL) in patients less than 163 cm in height. MEASUREMENTS AND RESULTS: Demographic and clinical data on all patients showed no change after initial interventions, followed by significant drop (8 to 2%) in incidence of balloon leak when smaller, shorter balloons were placed in shorter patients. CONCLUSIONS: Placement of larger, longer balloons in patients increases risk of perforation of balloon by calcific plaque in the distal thoracic and abdominal aorta.
Medical subject headings
- Intra-Aortic Balloon Pumping