Evaluation of laboratory methods for cystic fibrosis carrier screening: reliability, sensitivity, specificity, and costs.
Level III
Where this comes from
- Record sourced from PubMed, PMID 7525964.
- Also identified by PMC identifier 1049977.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report a comparative evaluation of three different laboratory methods for screening large numbers of mouthwash DNA samples for common cystic fibrosis mutations. Sensitivity, specificity, and costs of ARMS (allele refractory mutation detection system), dot blotting, and a deletion/digest/PAGE method (multiplex PCR of exons 10 and 11, digest with HincII followed by polyacrylamide gel electrophoresis (PAGE)) were assessed. ARMS was the most reliable and sensitive method and so was considered more suitable than the cheaper deletion/digest/PAGE. As well as being less reliable than ARMS, the dot blotting method assessed was considerably more costly. ARMS was the best laboratory method for CF screening tested.
Medical subject headings
- Cystic Fibrosis
- DNA Mutational Analysis
- Genetic Testing