Spiral CT of the chest: comparison of cine and film-based viewing.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 7568857.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine radiologists' ability to find lung nodules on spiral computed tomographic (CT) scans of the chest with both rapid sequential (cine) and conventional film-based viewing. Eight radiologists searched for lung nodules on spiral CT images (10-mm collimation, 10 mm/sec table speed) presented in two formats. Cine viewing was performed at a computer work-station; sections were viewed in 2-mm increments at frame rates up to 10 frames per second. Film-based viewing of images from a laser printer was performed with a lightbox; sections were viewed at 4-mm increments. Eight 3-5-mm-diameter simulated nodules were superimposed on each of five normal CT scans. Radiologists found a higher fraction of nodules with the cine presentation than with film (mean, 0.69 +/- 0.02 [standard error] versus 0.58 +/- 0.03, respectively [P = .006]). Diameter thresholds for nodule detection (50% correctly localized) were 3.3 and 3.5 mm, respectively. Cine viewing of spiral CT images of the chest improved radiologists' ability to detect nodules.
Medical subject headings
- Solitary Pulmonary Nodule
- Tomography, X-Ray Computed