Arterial blood sampling devices influence ionized calcium measurements.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 7587257.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine the effect of commercially available arterial blood sampling devices on ionized calcium measurements. Prospective study. Neurosurgical and shock-trauma intensive care units (ICU) at a tertiary care teaching hospital. Fourteen patients admitted to the ICU. Each patient had an indwelling arterial catheter. Arterial blood sampling. In 14 ICU patients, measurements of arterial blood ionized calcium concentrations were performed, using 12 different commercially available arterial blood sampling devices. Significant underestimation of ionized calcium in blood samples compared with the reference test tube (Vacutainer 45) was seen in seven of the devices. Arterial blood ionized calcium concentrations measured, using one commercially available syringe, were significantly higher compared with the reference test tube. There was no correlation between either the amount or type of heparin in the arterial blood sampling devices and arterial blood ionized calcium measurement. This study demonstrates that various commercially available arterial blood sampling devices alter arterial blood ionized calcium measurements. These alterations are clinically important because ICU patients may be treated with inappropriate calcium supplementation.
Medical subject headings
- Blood Chemical Analysis
- Calcium