SPECT imaging of fluorine-18.

Leichner, P K; Morgan, H T; Holdeman, K P; Harrison, K A; Valentino, F; Lexa, R; Kelly, R F; Hawkins, W G et al. · J Nucl Med · 1995

basic_science · Level V

Where this comes from

Abstract

The objective of this work was to determine the potential clinical usefulness of SPECT to image 511-keV annihilation photons. A triple-headed gamma camera equipped with ultra-high-energy collimators was used to image 18F. Sensitivity measurements were carried out and the FWHM and FWTM were determined in air and for a unit-density scattering medium. Additionally, tomographic phantom studies were acquired to evaluate image quality. The sensitivities of the three cameras were, for all practical purposes, identical. At a source-to-collimator distance of 100 mm, the FWHM and FWTM were 13 and 29 mm, respectively. A tomographic phantom study demonstrated that spheres with a diameter of 20 mm were well resolved when filled with 18F activity and placed inside a water-filled phantom. The triple-headed SPECT camera in this investigation is a practical means of acquiring tomographic 18F images. The reconstructed slices were of sufficient quality to be of value in some clinical studies.