Correlation of breast dose heterogeneity with breast size using 3D CT planning and dose-volume histograms.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 7631027.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of this study was to evaluate the effect of breast size on dose heterogeneity. Twenty women underwent a planning CT scan of the thorax. A three-dimensional treatment plan was devised for each patient using a standard technique of isocentric medial and lateral wedged tangential fields. Three-dimensional dose distributions were derived using an equivalent path length (EPL) inhomogeneity correction and cumulative dose-volume histogram (DVH) data calculated for the breast. Analysis of the DVHs for each patient reveals that 0.2-23.8% of the breast received an absorbed dose outside the desired 95-105% of that prescribed at the isocentre. The degree of dose heterogeneity was most strongly correlated with breast volume (r = 0.70, 95% confidence interval (C.I.) 0.37-0.87). There was also a positive correlation for breast dose heterogeneity versus brassière (bra) cup size (Spearman rank correlation rho = 0.62), breast area (r = 0.39, 95% C.I. -0.06-0.71) and chest wall separation (r = 0.31, 95% C.I. -0.15-0.66). We conclude that breast size is an important determinant of dose heterogeneity within the breast.
Medical subject headings
- Breast Neoplasms
- Image Processing, Computer-Assisted
- Radiotherapy Planning, Computer-Assisted
- Tomography, X-Ray Computed