Force-sensing microprobe for precise stimulation of mechanosensitive tissues.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 7642187.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Quantitative study of the transduction mechanisms in mechanically sensitive nerve terminals has been impeded by the lack of instrumentation with which to generate precisely controlled, physically localized mechanical stimuli. We have developed high-resolution force sensing mechanical microprobes for use in the characterization of such nerve terminals. This paper describes their design, fabrication, and testing. A microprobe is comprised of a 0.5- to 2-mm long silicon cantilever beam projecting from a larger supporting silicon substrate. Acting as the variable leg of a Wheatstone bridge circuit, a piezoresistive polysilicon element located at the base of the beam is used to measure the stimulation force applied at the tip. The microprobes exhibit a stable, linear relationship between the stimulation force and the resulting output voltage signal. Stimulation forces up to 3 mN have been generated with a measurement resolution of 10 microN. These microprobes have been used as the force sensing element of a closed loop feedback-controlled stimulation system capable of stimulating the mechanoreceptive nerve terminals of the rabbit corneal epithelium.
Medical subject headings
- Mechanoreceptors
- Physical Stimulation