Swanson silicone finger joint implants: a review of the literature regarding long-term complications.
review · Level V
Where this comes from
- Record sourced from PubMed, PMID 7642924.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A review of the English-language literature pertaining to complications associated with Swanson Silastic (Dow Corning, Midland, MI) finger implants was performed. The review focused on seven broad categories of complications: synovitis, lymphadenopathy, bone change, implant fracture, implant loosening, infection, and implant removal. Outcome data on 15,556 small joint implants were identified in 70 pertinent articles. The prevalence rates of complications associated with finger implants were all very low. Implant fracture and bone changes occurred in 2% and 4% (respectively) of finger implants. Each of the other complications developed in 1% or less of implants, including particulate synovitis (0.06% of implants). Particle-wear lymphadenopathy was reported in 0.08% of implants. The literature review identified no reports of immunologic reactions, connective tissue disease, or other systemic effects associated with the use of Swanson finger implants.
Medical subject headings
- Finger Joint
- Joint Prosthesis
- Prostheses and Implants
- Silicone Elastomers
Anatomy
- hand