Automatic patient-instruction devices in thin-section CT of the thorax: impact on image quality.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 7644653.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine what influence automatic patient-instruction (API) devices have on image quality of chest computed tomographic (CT) scans and whether the qualitative outcome justifies their routine use. Thin-collimation CT scans of two age- and sex-matched groups of 64 patients each were evaluated prospectively for the presence of breathing artifacts and for concomitant deterioration of image quality. Breathing commands in group 1 were given with the API device and in group 2 with technologist-performed patient instruction. Cardiac motion artifacts were not evaluated. The frequency of scans repeated owing to breathing artifact was determined. Image quality was worse in group 1 compared with that of group 2. The percentage of scans repeated was higher with API (38%) than without API (16%). API devices cannot be recommended for thin-section CT of the thorax. The large number of scans that must be repeated leads to a considerable increase in patient irradiation, scanning time, and cost.
Medical subject headings
- Computer-Assisted Instruction
- Patient Education as Topic
- Radiographic Image Enhancement
- Radiography, Thoracic
- Tomography, X-Ray Computed