The estimation of inspired isoflurane concentration in a low-flow system.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 7645702.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We have examined the predictability of inspired isoflurane concentration during low-flow anaesthesia using a to-and-fro breathing system. Twenty one adult patients requiring mechanical ventilation of the lungs during surgery took part in this study. Using a fresh gas flow of 2 l.min-1, the ratio of inspired isoflurane concentration to isoflurane vaporizer setting was found to be approximately 4/5th after 5 min of anaesthesia. The ratio was maintained throughout the procedure, except for a few minutes following each change in vaporizer setting.