Predictive analog to digital conversion of Doppler ultrasound signals.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 7698781.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Analog to digital conversion in multigate Doppler ultrasound systems for blood velocity measurements is a technological challenge. The echoes must be digitized at a rate determined by the system bandwidth (typically 2 MHz), and the dynamic range is large (16 bits or more) due to the presence of strong, low-frequency Doppler clutter echoes originating from slowly moving tissue. Off-the-shelf A/D converters do not meet these requirements with the transducer configuration employed by contemporary Doppler systems. Analysis reveals a 5-b reduction in required wordlength for an A/D converter in a predictive feedback loop when the maximum clutter frequency is about 1.5% of the pulse repetition frequency. The prediction error filter is recursive. Alternatively, first- and second-order DPCM (Differential Pulse Code Modulation) yield 4 and 6 b, respectively. With short input segments (from a high-resolution Color Flow Mapper), the results are, in the above order, 4, 4, and 5 b. The results are verified by processing an experimental Doppler signal.
Medical subject headings
- Analog-Digital Conversion
- Signal Processing, Computer-Assisted
- Ultrasonography, Doppler