A comparison of three variable performance devices for postoperative oxygen therapy.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 7710024.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Eleven patients in the intensive care unit following major abdominal surgery with a nasogastric tube in situ and receiving oxygen via facemask were allocated to receive in a random sequence oxygen at 4 l.min-1 via a Hudson mask, nasal cannulae or a nasal catheter with foam collar at the distal end. A significantly greater PaO2 was achieved using both the nasal catheter with foam collar (p < 0.01) and Hudson mask (p < 0.05) compared to the nasal cannulae. There were no significant differences in the PaCO2 or SpO2 values obtained between the devices. In the majority of postoperative patients in whom a variable performance device is indicated, nasal devices appear preferable in terms of patient comfort and compliance. The nasal catheter with foam collar produced a significantly greater PaO2 than nasal cannulae in patients with a nasogastric tube in situ.
Medical subject headings
- Catheterization, Peripheral
- Masks
- Oxygen
- Oxygen Inhalation Therapy
- Postoperative Care