An investigation of the bending stiffness of and the plane stresses generated by a flanged external fixator pin.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 7714660.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In an attempt to reduce pin loosening, a flanged external fixator pin has been designed and its bending stiffness has been compared with that of a standard pin. The pins were inserted into pilot holes previously drilled into a piece of teak hardwood. Loads of different magnitudes were applied at a fixed moment arm and force-deflection curves were obtained. Thereafter, percentage stiffness increase was calculated for each pilot hole size. The results show that the addition of a collar to the external fixator pin increases its stiffness and its ability to resist bending forces. In a parallel study, the stresses generated at the pin-bone interface by this pin and a standard pin were compared using finite element analysis techniques. The results show that the flange significantly reduced the stresses generated at the pin-bone interface. In addition, stresses were dissipated over a wider area.
Medical subject headings
- Bone Nails
- External Fixators