An investigation of the bending stiffness of and the plane stresses generated by a flanged external fixator pin.

Oni, O O; Capper, M; Soutis, C · J Orthop Trauma · 1995

biomechanical · Level V

Where this comes from

Abstract

In an attempt to reduce pin loosening, a flanged external fixator pin has been designed and its bending stiffness has been compared with that of a standard pin. The pins were inserted into pilot holes previously drilled into a piece of teak hardwood. Loads of different magnitudes were applied at a fixed moment arm and force-deflection curves were obtained. Thereafter, percentage stiffness increase was calculated for each pilot hole size. The results show that the addition of a collar to the external fixator pin increases its stiffness and its ability to resist bending forces. In a parallel study, the stresses generated at the pin-bone interface by this pin and a standard pin were compared using finite element analysis techniques. The results show that the flange significantly reduced the stresses generated at the pin-bone interface. In addition, stresses were dissipated over a wider area.