External fixation of upper limb fractures: the effect of pin offset on fixator stability.

Oni, O O; Capper, M; Soutis, C · Biomaterials · 1995

biomechanical · Level V

Where this comes from

Abstract

A Shearer fracture fixator was applied to a hardwood model using various pin configurations and load-displacement curves were obtained from compressive and bending loads. The results show that offsetting the pins by 60 degrees increases rigidity by a factor of 4-5.