Electron-beam CT: use of a calibration phantom to reduce variability in calcium quantitation.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 7784560.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To measure scanner and patient variation in computed tomographic (CT) numbers for electron-beam CT and to determine the ability of calibration phantoms to reduce variability in calcium quantitation. Two calibration phantoms were imaged to ensure longitudinal homogeneity and to determine the short-term intrascanner variation in CT numbers. Each phantom set was imaged twice a day for 14 weeks to determine intra- and interscanner variation. Data from examinations of 167 patients that included the phantom were analyzed to determine the intra- and interpatient variation in CT numbers of objects with known calcium concentrations. The calibration reduced scanner variations by approximately 25%. The calcium concentration associated with a CT number of 130 HU varied from 77.1 to 136.4 mg/cm3 and was dependent on patient girth, sex, smoking history, and image level. Scanner and patient variations in CT numbers in electron-beam CT can be reduced with a calibration phantom. In vitro and in vivo estimates of calcium concentration had a precision of 2% and 7%, respectively.
Medical subject headings
- Calcinosis
- Coronary Disease
- Tomography, X-Ray Computed