Measurement of tidal flow using a transit-time ultrasonic breath analyser.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 7793550.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The ability of the Transit-time Ultrasonic Breath Analyser (TUBA, GHG Medical Electronics GMBH, Zürich, Switzerland) to measure peak flow and tidal volume in the laboratory was tested using a variety of flow and pressure conditions, chosen to simulate the respiratory patterns of patients receiving mechanical ventilatory support. A stable zero baseline was achieved by acoustic damping of the TUBA flow sensor head. A piston pump was used to generate sinusoidal flow pattern, with a peak flow range from 0.1 to 1.51.s-1. The calculated peak flow matched the peak flow measured by the TUBA. The TUBA accurately measured tidal volumes (+/- 10%) delivered using three different flow patterns over a range of volumes from 0.25 to 11. We conclude, that once modified, the TUBA can provide an accurate measurement of peak flow and tidal volume over a range of values likely to be encountered during mechanical ventilation of the lungs.
Medical subject headings
- Breath Tests
- Peak Expiratory Flow Rate
- Tidal Volume
- Ultrasonography