The need for quality control in measurement of mixed venous oxygen saturation.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 7802188.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We investigated the variability of arterial, mixed venous and peripheral venous oxygen saturation readings in three co-oximeters in regular use on two Intensive Care Units. Over a 2-week period, 96 readings were obtained from each of the following machines; two AVL 912 co-oxylite co-oximeters (Machines A and B, Biomedical Instruments, Graz, Austria) and one IL-282 (Machine C, Instrumentation Laboratories, Lexington, MA, USA). The coefficient of variation in the arterial oxygen saturations was extremely small on all three machines (< 0.43%). The coefficient of variation in the mixed venous samples, however, ranged from 1.96 to 4.61% on machine A, and from 2.73 to 4.71% on machine B, but only from 0.17 to 1.47% on machine C. The variation in mixed venous saturations obtained from a single blood sample repeatedly analysed on machines A and B was large enough to influence clinical management.
Medical subject headings
- Oximetry
- Oxygen